MA 200/1.X2.IC.A.NS Moisture Analyzer with Automatic Drying Chamber - High Precision Laboratory Instrument

PreviousNext
image-brand

Product order number: WL-306-0079

24.14684 RON

28.73473 RON

(VAT included)
  • Delivery in 2-4 weeks
  • Plată cu card online sau cu OP
  • 021 255 3132

Description

MA 200/1.X2.IC.A.NS Moisture Analyzer MA X2.IC.A moisture analyzers are hi-tech measuring instruments intended for quick determination of relative moisture content, dry mass content and other parameters in samples of different substances.

MA.X2.IC.A series is equipped with innovative system: the drying chamber can be opened and closed automatically using button or proximity sensors.

Such solution allows:

- Maintaining moisture analyzer clean – operator does not touch moisture analyzer’s housing.

- Removing any shocks caused by manual closing of the chamber – the chamber closes automatically and always with the same intensity.

- Achieving high repeatability of opening and closing the drying chamber. - Enhancing drying process automation.

- Enhancing safety of operation by eliminating contact with hot components of drying chamber. - Improving operation ergonomics.

MA.X2.IC.A series features user-friendly colour touchscreen. Free configuration of the control panel can be done by means of hotkeys, customizable information fields and programmable labels. Drying process can be carried out for any temperature and – thanks to databases – particular drying process may be set for a specific product. Highlights of X2 Series moisture analyzers: - Automatic opening and closing of the drying chamber - 5” capacitive colour touchpad - Modern interface - Uncomplicated, intuitive operation - Programmable display - Drying profiles (standard, mild, step, fast) - GLP/GMP printouts-reports - Repeatable and customized applications - Work optimization due to the use of halogen lamps Capacity and temperature MA.X2.IC.A maximum capacity ranges from 50g/0.1mg to 210g/0.1mg.

Moisture content is measured with the 0.01% accuracy (0.001 % for samples bigger than 1.5g). The maximum drying temperature is 160˚C (there is an optional moisture analyzer with 250˚C maximum temperature). An additional element for moisture analyzers that improves their functionality is a water vapour permeability determination set used in tanning industry for determining hygienic properties of leather. Databases: Information system of X2 series moisture analyzers is based on 8 databases, allowing many users to operate product database comprising many samples.

Collected measurements may be subjected to subsequent analysis.

Collected data is registered in 8 databases:

- Products (up to 5000 products)

- Users (up to 100 users)

- Packaging (up to 100 packagings)

- Customers (up to 1000 customers)

- Drying programs (up to 200 drying programs)

- Drying process reports (up to 5000 drying process reports)

- Ambient conditions (up to 10000 records)

- Weighings (up to 50000 records)

The results

The results obtained using the moisture analyzer (the final result of the drying process):

- % M – the mass loss in %; displays the mass change recorded during the drying process, expressed as a percentage,

- % D – the dry mass obtained during the drying process, expressed as a percentage; the result is the portion of the sample that remains on the weighing pan after the moisture has evaporated,

- % R – the ratio of moisture content to dry mass, obtained during the drying process, expressed as a percentage; the result is the portion of the sample that evaporated during the drying process, - g – the sample mass after drying. Interfaces USB interface ensures quick transfer and copying of any results of your work (measurements, reports, databases) to a computer or other moisture analyzer. Remote access to the moisture analyzer allows for online data management. Cooperation with E2R allows for external data management which in turn increases the efficiency and results to the maximum.
 

Metrological parameters  
Maximum capacity [Max] 200 g
Readability [d] 0.1 mg
Tare range -200 g
Adjustment internal (automatic)
Maximum sample weight 200 g
Heating module metal heater
Moisture content repeatability +/-0.05% (sample ~ 2g), +/-0.01% (sample ~ 10g)
Moisture content readability 0.0001%
Drying temperature range Max 160 °C
Physical parameters  
Leveling system manual
Display 5” graphic color touchscreen
Weighing chamber automatic
Maximum sample height 20 mm
Weighing pan dimensions ø90, h= 8 mm
Packaging dimensions 476×381×346 mm
Net weight 5.3 kg
Gross weight 6.8 kg
Construction  
Protection class IP 43
Features of use  
Display mode %MC, %DC, %ATRO MC, %ATRO DC, g, g/kg MC, g/kg DC, -%MC
Drying method 4 drying profiles (standard, fast, step, mild)
Finish mode 4 modes (automatic, manual, time-defined, user-defined)
Additional functions sample traceability
Communication interface  
Communication interface RS232¹, USB-A, USB-B, Ethernet, Wi-Fi®
Electrical parameters  
Power supply 100V – 120V AC 50/60Hz or 200V-240V AC 50/60Hz
Power consumption 6 W
Heating module power 450 W
Environmental conditions  
Operating temperature +10 — +40 °C
SOP – metrological tests  
Mass indication control External mass standard 50 g, 100 g, 200 g (depending on the type of moisture analyzer)
Internal adjustment For .IC models only, GLP report
External adjustment For all models, GLP report
Drying temperature control One- or two-point, depending on the drying profile (equipment: GT105k-12/Z Control Thermometer)
Moisture content indication control Certified MV Test Set
Adjustment Internal / External adjustment Depending on the type of moisture analyzer, GLP report
Drying temperature adjustment Three-point adjustment at defined temperatures, programmable values (equipment: GT105k-12/Z Control Thermometer)